Description
M-740 AC/DC Clamp Milliammeter (CE)
The M-740 AC/DC Clamp Milliammeter (CE) is a leakage current measuring instrument designed for measuring low-level AC and DC currents using a 40 mm clamp current transformer (CT). It provides DC measurement ranges of 100 mA and 1000 mA, along with AC measurement ranges of 100 mA, 1000 mA, and 10 A.
The M-740 offers a minimum resolution of 0.01 mA for low-current measurements and features zero adjustment for DC current measurement. Its separate clamp CT allows current measurement without disconnecting the conductor, while the DC analog output of 100 mV full scale on each range enables connection to external recording equipment. The instrument also supports data hold and automatic power-off, with safety ratings of CAT II 600 V and CAT III 300 V.
Accuracy Specifications
Conditions: 23°C ± 5°C, less than 85% RH.
| Function | Range | Measuring Range | Resolution | Accuracy |
|---|---|---|---|---|
| DC Current (DCA) | 100 mA | 0.1 to ±99.99 mA | 0.01 mA | ±1% rdg ± 10 dgt |
| 1000 mA | 1 to ±300 mA | 0.1 mA | ±1% rdg ± 10 dgt | |
| ±300 to ±700 mA | 0.1 mA | ±2% rdg ± 10 dgt | ||
| ±700 to ±999.9 mA | 0.1 mA | ±3% rdg ± 10 dgt | ||
| AC Current (ACA), 50/60 Hz | 100 mA | 0 to 99.99 mA | 0.01 mA | ±1% rdg ± 10 dgt |
| 1000 mA | 0 to 999.9 mA | 0.1 mA | ±1% rdg ± 10 dgt | |
| 10 A | 0 to 9.999 A | 0.001 A | ±1% rdg ± 10 dgt |
Additional Specifications:
- Influence of Terrestrial Magnetism: Less than ±2.0 mA
- Influence of Magnetization: Less than ±2.0 mA by DC 1.5 A on/off
- Maximum Input Current (DC): 1.5 A
- Maximum Input Current (AC): 20 A
Key Features –M-740 AC/DC Clamp Milliammeter (CE)
-
AC/DC Leakage Current Measurement: DC ranges of 100 mA and 1000 mA; AC ranges of 100 mA, 1000 mA, and 10 A (45–65 Hz).
-
High Measurement Resolution: Minimum resolution of 0.01 mA for low-level leakage current measurement.
-
40 mm Clamp CT: Separate clamp-type current transformer with a 40 mm jaw opening.
-
DC Analog Signal Output: 100 mV full-scale output on each measurement range for connection to external recording equipment.
-
Safety Rating: IEC 61010-1 / IEC 61010-2-032, CAT II 600 V or CAT III 300 V.
Documents
Catalog: e_ca_m740.pdf
Instruction Manual: e_ma_m-740.pdf
Additional Information
Measuring method: |
Clamp CT Type |
AC Conversion: |
Dual integration method |
Accuracy: |
See accuracy specifications in description tab |
Display: |
Max. 2000 count on LCD with annunciator |
Current Consumption: |
Approx. 9mA (approx. 200h for continuous use) |
Option: |
Cable for recorder |
Jaw opening capability: |
40mmφ |
Over range indication: |
"OL" mark on LCD |
Data hold indication: |
"DH" mark on LCD |
Sampling rate: |
1 time/sec. for DC and 6 times/sec. for AC |
Withstanding Voltage: |
AC 3700V/1 minute between CT and outer case |
Storage Temperature: |
-10~60℃, < 70%RH (without condensation) |
Low battery indication: |
"B" mark on LCD |
Auto power off: |
Approx. 10 minutes after power on and can absolve this function by switch |
Power supply: |
UM-4×4 |
Battery life: |
Approx. 60 hours continuous |
Size: |
Display Part 78(W)×155(H)×32(D)mm, approx. 280g || CT (CTP-40DC) 69(W)×175(H)×24(D)mm, approx. 180g |
Accessories: |
Soft Case, Batteries, Instruction Manual |
Limitation of circuit voltage: |
less than AC/DC 600V |
Measuring function: |
AC/DC current |
Zero adjustment: |
For DC current range, by "0 ADJ" switch |
Signal Output: |
DC 100mV full scale to each range (output impedance: less than 10KΩ) |